Assessing cassette printer reliability.
Rigorously tested for your peace of mind.
StatLab Cassette Printers tested for equivalent use of 24 years.
Background
StatLab is committed to designing and producing highquality, best-in-class printers which reduce the burden of maintenance and attention needed throughout the product lifetime, enabling laboratories to spend more of their time on patient care. During development of the PiSmart cassette printers, accelerated life testing was performed to ensure the product delivered on high performance expectations. We are proud to say that all production units were shipped with this level of quality built-In!
StatLab is committed to designing and producing highquality, best-in-class printers which reduce the burden of maintenance and attention needed throughout the product lifetime, enabling laboratories to spend more of their time on patient care. During development of the PiSmart cassette printers, accelerated life testing was performed to ensure the product delivered on high performance expectations. We are proud to say that all production units were shipped with this level of quality built-In!
Methodology
A test was designed to measure the reliability of StatLab PiSmart cassette printers over the typical lifetime of the instrument. Typical tests like these are accelerated in order to complete a typical instrument operation in a much shorter period of time. Five cassette printers were used for the testing, each run with a single cycle to include:
A test was designed to measure the reliability of StatLab PiSmart cassette printers over the typical lifetime of the instrument. Typical tests like these are accelerated in order to complete a typical instrument operation in a much shorter period of time. Five cassette printers were used for the testing, each run with a single cycle to include:
- A full movement of the dispense from home to eject of cassette and return to Home.
- Rotation of the clamp assembly from Home to Clamp, pause, move from Clamp to Home.
- Movement of the print assembly from home position to eject position and return to home position.
Accelerated Life Test Results1
At the completion of the testing, the units were inspected and no wear to the cassette delivery or clamp assembly was seen. Two printers were run with an additional 310,000 cycles. The units were inspected and no wear to the slide delivery or clamp assembly was seen.
At the completion of the testing, the units were inspected and no wear to the cassette delivery or clamp assembly was seen. Two printers were run with an additional 310,000 cycles. The units were inspected and no wear to the slide delivery or clamp assembly was seen.
Conclusion
The five units were run beyond the planned 300,000 cycles with two of these units running an additional 310,000 cycles for a total of over 639,000 cycles on each instrument. This is an equivalent of approximately 24.58 years of typical operation on each printer. No problems were reported. This test of five printers exceeded the planned 300,000 cycle test to sufficiently ensure runtime test requirements equivalent to ten years were met for the entire study, with two printers running the equivalent of an additional 14 years each for a total of 24.58 years of average use.This test is considered to not only be a pass, far exceeding the expected lifetime of laser and thermal transfer cassette printers. The result demonstrates a highest level quality of product that is ready for use in any anatomic pathology lab setting.
The five units were run beyond the planned 300,000 cycles with two of these units running an additional 310,000 cycles for a total of over 639,000 cycles on each instrument. This is an equivalent of approximately 24.58 years of typical operation on each printer. No problems were reported. This test of five printers exceeded the planned 300,000 cycle test to sufficiently ensure runtime test requirements equivalent to ten years were met for the entire study, with two printers running the equivalent of an additional 14 years each for a total of 24.58 years of average use.This test is considered to not only be a pass, far exceeding the expected lifetime of laser and thermal transfer cassette printers. The result demonstrates a highest level quality of product that is ready for use in any anatomic pathology lab setting.
Figure 1. Accelerated Life Testing Cycles
Five printers were evaluated for one year or more of typical use. Each horizontal line in the chart represents one year of typical use.
Five printers were evaluated for one year or more of typical use. Each horizontal line in the chart represents one year of typical use.
REFERENCES
1. PI007-D087 Reliability Test Report v3
1. PI007-D087 Reliability Test Report v3
Recent Posts
-
Why HCA Mid-America Division only uses KT slides in their lab.
One lab's story of overcoming tissue wash challenges.
-
Inspection Coming! 10 steps to help prepare for CAP/CLIA inspections.
Is inspection season approaching and you’re starting to feel that familiar mix of urgency and uncert
-
Why I Love Histology: Celebrating the Art and Science Behind Every Slide
Click to watch the Why I Love Histology Histohints Webinar.